82 Complete Function Check
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
26
Adjust the set rate to 5 BPM.
27
The alarm shall disappear in 30 seconds.
Check the Low Breath Rate Alarm
28
Adjust the ventilator settings as follows:
29
The low breath rate alarm shall occur within 30 seconds.
30
Adjust the set rate to 20 BPM.
31
The alarm shall disappear in 30 seconds.
Check the Low Volume Alarm
32
Adjust the ventilator settings as follows:
33
The low volume alarm shall occur within 20 seconds.
34
Adjust the low volume alarm to 0.05 l.
35
The alarm shall disappear in 30 seconds.
Check the Switching of Power Source to the Internal Battery
36
Disconnect the power cord from the AC power supply.
37
Treatment continues and the Vivo 40 runs on internal battery with the
internal battery LED lit.
38
Reconnect the power cord to the AC power supply.
39
The ventilator shall switch to main supply with the AC LED lit.
S
ETTING
V
ALUE
Set rate
5 BPM
High breath rate alarm
Off
Low breath rate alarm
15 BPM
S
ETTING
V
ALUE
High breath rate alarm
Off
Low volume alarm
2.0 l