Using the Vivo 40
Vivo 40 clinician’s manual
Doc. 003886 En-Us A-1e
The following are examples of the operation time with new, fully charged bat-
teries and Vivo 40 running in PCV mode:
External Battery
The Vivo 40 can be operated from a 12 V or a 24 V DC external battery.
• Use the battery cable 12/24 V DC and check carefully that the voltage is
12 V or 24 V.
• Check the polarity of the external battery before connecting it to the
Vivo 40.
With an external battery connected, the Vivo 40 will automatically switch over
to the external battery source if the AC power cord is removed or if the AC
power supply fails. The external DC level is shown under “Others, Device infor-
mation” in the menu.
S
ETTING
V
ALUES
EPAP
4
8
4
IPAP
10
20
30
Breath Rate
12
15
20
Insp. Time
2.0
1.5
1.5
Rise Time
3
3
1
Insp. Trigger
Off
Off
Off
Delivered Tidal
Volume
0.20
0.50
0.75
Total Time
4 h
2 h 45 min
1 h 45 min
E
NVIRONMENTAL
C
ONDITIONS
Ambient temperature
68°F (20°C)
Altitude
Sea level