OPERATION MANUAL

OM-ER436E

Cordless Handpiece

Summary of Contents for EndoSequence II

Page 1: ...OPERATION MANUAL OM ER436E Cordless Handpiece...

Page 2: ...l Do not use this for implants other than endodontic treatment or other dental treatment Guidance and manufacturer s declaration electromagnetic emissions The EndoSequence II is intended for use in th...

Page 3: ...les 70 Ut 30 dip in Ut for 25 cycles 5 Ut 95 dip in Ut for 5 secs 3 A m Compliance level 6kV contact 8kV air 2kV for power supply lines 1kV line to line 2kV lines to earth 5 Ut 95 dip in Ut for 0 5 cy...

Page 4: ...ld be observed to verity normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the EndoSequence II b Over the frequency range 150k...

Page 5: ...ers rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the...

Page 6: ...product properly These indicators are to allow you to use the product safely and prevent danger and harm to you and others These are classified by degree of danger damage and seriousness All indicator...

Page 7: ...es Do not drop the motor handpiece or battery charger Place the battery charger on a flat and stable surface Should the battery fluid leak and get into your eyes immediately wash eyes thoroughly with...

Page 8: ...rce Before changing the head or file turn off the power of the motor handpiece Changing with the power on may cause unintended rotation by accidental activation of the ON OFF switch When inserting the...

Page 9: ...and operating time become significantly shorter since the charging current differs The motor handpiece consumes electricity very slightly even when the power is off In addition fully charged rechargea...

Page 10: ...tion The built in feedback circuit which keeps rotation at a constant speed even when the load on the motor handpiece changes Non contact charger prevents improper charging due to deterioration of met...

Page 11: ...Pressing this switch when the power is on rotates the motor handpiece and de pressing the switch again stops the handpiece Normal rotation operation Pressing and holding the switch for approx one seco...

Page 12: ...rque suitable for contra angle which you use For gear ratio 4 1 10 1 16 1 or 20 1 can be set LCD Panel AUTO REVERSE The current auto reverse mode mark is displayed AUTO REVERSE If the load is removed...

Page 13: ...are drained or in a remarkably low voltage Charge the batteries ALARM Symbol The present alarm symbol is displayed Alarm ON Alarm OFF No display The symbol indicating the remaining amount of the batt...

Page 14: ...this time check that the power lamp lights up Fig 5 4 Insert the motor handpiece into the Battery Charger Charging starts with the charge mark flashing on LCD screen 5 When the alarm sounds and is dis...

Page 15: ...s charging cannot be properly completed and the battery compartment may feel hot We recommend that you charge batteries for longer periods if possible The power of the motor handpiece should not be tu...

Page 16: ...it engages with the latch mechanism Push it inward to click Fig 7 File Removal depress the push key and pull out the file Fig 8 When attaching and detaching the file turn off the power beforehand Aft...

Page 17: ...or handpiece starts while the key is pressed If you release the key it stops Intermittent operation Auto Reverse Function Auto reverse setting can be changed by pressing the SELECT key for one second...

Page 18: ...ad higher than the set torque limit value AUTO REVERSE The handpiece starts in forward rotation When a load higher than the torque limit is applied the file will rotate in reverse When the load is rem...

Page 19: ...Calibration This function is to decrease fluctuation in the rotation speed of the motor handpiece and the difference in torque by the contra angle head 1 Lubricate the contra angle head with BRASSELER...

Page 20: ...5 Display on the liquid crystal panel LCD returns to the original state if no operation is performed for a while This function does not work unless remaining battery capacity is sufficient Perform ca...

Page 21: ...and it may flows backward 2 Cleaning Motor Handpiece When the motor handpiece becomes dirty wipe it off with a cotton cloth moistened with rubbing alcohol 3 Cleaning Charger When the charger becomes d...

Page 22: ...in at 121 C 250 F or 15 min at 132 C 270 F 5 Keep the handpiece in the autoclave pouch to keep it clean until you use it Sterilization at 121 C for more than 15 minutes is recommended by EN ISO 17665...

Page 23: ...sulting from you changing the batteries yourself and not following the CAUTIONS ON CHANGING BATTERIES 11 Changing Batteries Do not open any part other than the battery cover Use only batteries as spec...

Page 24: ...e old batteries 6 Insert new batteries according to the polarity marking in the battery box A mix up between the plus and minus sides will not allow operation of the handpiece 7 Attach the battery cov...

Page 25: ...on of the circuit may be suspected Contact Brasseler USA Replace the motor handpiece or contra angle head Malfunction of circuit The motor handpiece is locked at the time of the auto reverse mode Malf...

Page 26: ...seler USA Check that no battery other than Ni MH battery is used Correctly set Remove metal on the charger See 12 Error Code Insert the power cord plug into the outlet Insert the power cord jack into...

Page 27: ...ge the batteries Charge the batteries Charge the batteries Use in a warm room Clean the contra angle head If nothing is displayed on the liquid crystal panel of the handpiece even when it is set in th...

Page 28: ......

Page 29: ...08 00 00 N ByYour Side in Dentistry One Brasseler Boulevard Savannah GA 31419 To order call 800 841 4522 or fax 888 610 1937 Visit our website www BrasselerUSA com...