BpTRU Operator’s Manual
74
Temperature Celsius/Fahrenheit Mode (BPM300)
The BpTRU can be configured to display temperature
units in either Celsius or Fahrenheit. The default display
unit is Celsius, but the unit can be toggled as follows.
1.
Turn the BpTRU off using the power switch.
2.
To toggle temperature display units, hold
Mode
while
turning the BpTRU on again.
Repeating this toggles the BpTRU to the previous state.
When a temperature probe is attached, the selected units
(°C or °F) are illuminated on the BpTRU display.
Temperature Calibration Check (BPM300)
The Temperature Calibration Check allows a comparison
of the accuracy of the BpTRU’s temperature measurement
circuitry to a reference thermometer of known accuracy.
(A precision electronic thermometer is recommended, but
a mercury thermometer will work as well.)
1.
Attach a temperature probe.
2.
Turn the BpTRU off using the power switch.
3.
To select the Temperature Calibration Check mode,
hold
BP Start
while turning the BpTRU on.
4.
Release
BP Start
.
(Temperature Calibration Check mode is indicated by the
software version being displayed in the
Systolic
display).
The probe temperature is displayed (in °C or °F) as whole
degrees in the
Temperature
display, and hundredths of a
degree in the
Cycle
display, as shown below:
Firmw are
Rev ision
Degrees
(1/100
ths
)
Summary of Contents for BPM100
Page 2: ......
Page 14: ...Manuale di istruzioni BpTRU 8...
Page 20: ...BpTRU Operator s Manual 14...
Page 30: ...BpTRU Operator s Manual 24...
Page 48: ...BpTRU Operator s Manual 42...
Page 62: ...BpTRU Operator s Manual 56...
Page 82: ...BpTRU Operator s Manual 76...
Page 90: ...BpTRU Operator s Manual 84...