Guidance and manufacturer’s declaration – electromagnetic immunity
The BpTRU monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the BpTRU monitor should assure that it is used in
such an environment.
Immunity Test
IEC 60601 test
level
Compliance
level
Electromagnetic environment –
guidance
Conducted RF
IEC 6100046
Radiated RF
IEC 6100043
3 V rms,
150 kHz to 80
MHz
3 V/m,
80 MHz to 2.5
GHz
3 V rms
3 V/m
Recommended separation
distance:
d
1.2
P
=
d
1.2
P
=
(80 MHz to 800 MHz)
d
2.3
P
=
(800 MHz to 2.5 GHz)
where P is the maximum output
power rating of the transmitter
in Watts, according to the
transmitter manufacturer, and d
is the recommended separation
distance in metres.
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
a
,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: The guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Summary of Contents for BPM100
Page 2: ......
Page 14: ...Manuale di istruzioni BpTRU 8...
Page 20: ...BpTRU Operator s Manual 14...
Page 30: ...BpTRU Operator s Manual 24...
Page 48: ...BpTRU Operator s Manual 42...
Page 62: ...BpTRU Operator s Manual 56...
Page 82: ...BpTRU Operator s Manual 76...
Page 90: ...BpTRU Operator s Manual 84...