BpTRU Operator’s Manual
72
After approximately five seconds, the applied pressure is
displayed in whole mmHg in the
Pulse
display and in
hundredths of mmHg in the
Reading
display.
A properly calibrated monitor reads “0” in the
Pulse
display and less than “10” in the
Reading
display.
Return to normal operation by powering the BpTRU off
and then on again.
BP Reference Gauge Calibration Check
The BP Reference Gauge Calibration Check is a feature
that allows you to compare the accuracy of the BpTRU’s
pressure transducer against a reference pressure gauge
of known accuracy. (An electronic pressure gauge is
recommended, but a mercury or aneroid gauge is
acceptable.)
1.
Turn the BpTRU off using the power switch.
2.
Disconnect the coiled tubing from the BpTRU.
3.
To select the BP Reference Gauge Calibration Check
mode, hold
BP Start
while turning the BpTRU on.
4.
Release
BP Start
. The monitor now performs an
automatic zero-offset calibration.
BP Reference Gauge Calibration Check mode is indicated
by the software version being displayed on the
Systolic
display. The applied pressure is displayed in whole
mmHg in the
Pulse
display and in hundredths of mmHg
in the
Reading
display, as shown below:
Firmw are
Rev ision
Pressure
(1/100
ths
)
Pressure
Units
Summary of Contents for BPM100
Page 2: ......
Page 14: ...Manuale di istruzioni BpTRU 8...
Page 20: ...BpTRU Operator s Manual 14...
Page 30: ...BpTRU Operator s Manual 24...
Page 48: ...BpTRU Operator s Manual 42...
Page 62: ...BpTRU Operator s Manual 56...
Page 82: ...BpTRU Operator s Manual 76...
Page 90: ...BpTRU Operator s Manual 84...