BpTRU Operator’s Manual
22
The BpTRU will not begin a new measurement until the
cuff is fully deflated from the previous measurement.
Results of an Automatic Series of Measurements
The BpTRU performs the following actions at the
completion of six measurements.
A tone sounds (unless the tone has been disabled as
described in “Technical Modes of Operation” on
page 70).
The results of the last measurement are displayed for
five seconds, while the Reading display shows “6.”
After five seconds, the average of the last five
Systolic, Diastolic, and Pulse measurements is shown
in the respective display, while the Reading display
shows “A” (for average).
The first measurement is always discarded from the
average as it is often elevated due to patient stress
(white-coat effect). If an error code is displayed for any
measurement, that measurement is not included in the
average. The average can be computed from less than
five measurements.
Reviewing Measurements
Use the Review button to review previous
measurements.
Review Measurements in Single Mode
In Manual mode you can review previous manual
measurements by pressing the Review button. The
Review button cycles through all measurements taken
since Manual mode was invoked, or since the last time
the Clear button was pressed (up to a maximum of 99
previous measurements).
Pressing the
Review
button during a measurement has
no effect.
Summary of Contents for BPM100
Page 2: ......
Page 14: ...Manuale di istruzioni BpTRU 8...
Page 20: ...BpTRU Operator s Manual 14...
Page 30: ...BpTRU Operator s Manual 24...
Page 48: ...BpTRU Operator s Manual 42...
Page 62: ...BpTRU Operator s Manual 56...
Page 82: ...BpTRU Operator s Manual 76...
Page 90: ...BpTRU Operator s Manual 84...