16
WARNING
Prolonged, excessive, or incorrect use of the BpTRU
may lead to tissue injury, venous pooling, and
impairment of circulation in the arm of the patient.
If the low-battery indicator appears, a maximum of 20
blood pressure measurements may be taken before the
BpTRU prevents further operation. Plug the BpTRU into
a wall outlet to charge the battery. Note that it is still
possible to perform measurements while the battery is
charging.
To allow auto-calibration to complete, wait at least 5
seconds after turning on the monitor before pressing the
BP Start
button.
Obtaining Accurate BP Measurements
Careful patient treatment and use of the BpTRU will
result in more accurate blood pressure measurements.
Select an Appropriate Size of Cuff
The BpTRU may not provide accurate readings if you
use the wrong size of cuff. The five available cuff sizes
are listed below, with the range of arm circumferences
appropriate for each. The child and extra-large adult
sizes may be ordered separately.
Child
1318 cm
Small adult
1826 cm
Regular
adult
2634 cm
Large adult
3243 cm
Extra-large
adult
4152 cm