BpTRU Operator’s Manual
62
Item
Specification
Single temperature
measurement range
35 °C to 39 °C (95.0 °F to 102.2 °F)
Continuous temperature
measurement range
28 °C to 43 °C (82.4 °F to 109.4 °F)
Single temperature
measurement time
15 seconds
Continuous temperature
measurement settling time
3 – 4 minutes
Single temperature efficacy Average error /- 0.1 °C (0.2 °F)
Standard deviation within 0.4 °C (0.7 °F)
Continuous temperature
accuracy
Meets ASTM E 111200 accuracy
requirements
Input power requirements
to universal power supply
100 – 240 VAC, 390 mA, 50 – 60 Hz
Output power of universal
power supply
10 V, 1.0 - 1.25 A
Cord set connector type
IEC 320 C7 connector
Cord set plug type
North American, Continental Europe,
United Kingdom, Australia
Cord set length
2 m, 6.7 feet
Battery voltage and
capacity
8 V, 3.0 Ah
BP measurements per
battery charge
200 at 2 minute cycle time
Battery charge time
12 hours, trickle charge
Battery charge cycles
250
Environmental operating
ranges
Temperature: 10 to 40 degrees Celsius
Humidity: 15 to 90% RH (non-condensing)
Elevation: -170 to 1,700 m (from sea level)