Display Reference
53
markings, particularly the cuff model and the
markings for arm size range and index.
The temperature probes should be examined for
cracks, sharp edges, or other deformations that may
present a hazard to the patient or user.
Cleaning / Disinfection
The BpTRU may be cleaned as required by wiping
with a cloth dampened with warm water and a mild
detergent solution. Allow sufficient time to air dry.
The BpTRU and accessories, including the
temperature probe, may be disinfected as required
(for example, if bodily fluids come into contact the
monitor) by wiping with a cloth dampened with a
diluted, non-staining disinfectant solution such as
bacteriostatic detergent. Allow sufficient time to air
dry.
The coiled tubing and power supply cabling may be
cleaned by wiping with a cloth or sponge dampened
with warm water and a mild detergent solution. If
further cleaning is required, coiled tubing and cuffs
may be cleaned using rubbing alcohol. Do not use
harsh solvents such as acetone as they may destroy
these parts.
The blood pressure cuffs may be cleaned by wiping
with a damp cloth or sponge and a mild detergent
solution. Allow sufficient time to air dry.
CAUTION
Excessive wetting or immersion of the BpTRU and
accessories may lead to electrical safety hazards or to
blockage of connectors and tubing. Do not attempt to
use the monitor without allowing sufficient time to
air dry.
Summary of Contents for BPM100
Page 2: ......
Page 14: ...Manuale di istruzioni BpTRU 8...
Page 20: ...BpTRU Operator s Manual 14...
Page 30: ...BpTRU Operator s Manual 24...
Page 48: ...BpTRU Operator s Manual 42...
Page 62: ...BpTRU Operator s Manual 56...
Page 82: ...BpTRU Operator s Manual 76...
Page 90: ...BpTRU Operator s Manual 84...