BpTRU Operator’s Manual
32
Troubleshooting Guide
Symptoms
Possible Causes
Corrective Actions
Power is on (green
indicator light is on)
but
BP Start
button
does not initiate
inflation.
BpTRU
is
in
Automatic mode and
six readings have
already been taken.
Press
Clear
to clear
the memory and then
press
BP Start
to
initiate a new BP
measurement.
Cuff pressure from
previous measurement
not yet fully released.
Check pressure in
system if the problem
persists.
Perform a BP Zero
Calibration Check
(page 71) or BP
Reference Gauge
Calibration Check
(
page 72
)
.
Zero pressure
calibration not
complete.
Wait 5 seconds and
push the
BP Start
button again.
Power is on (green
indicator light is on),
but displays are all
blank.
BpTRU is in sleep
mode.
Press any button to
restore the display.
Power is on (green
indicator light is on)
and device
malfunction LED is
displayed ( ).
Unrecoverable system
error has occurred.
Turn power off and
ensure cuff is deflated.
Turn power on again.
Refer to
“Unrecoverable
System Errors” on
page 41.
Power is on (green
indicator light is on)
but there is only a dash
(-) in the left digit of
the Cycle display.
BpTRU is in hide
mode. Refer to
“
BP
Hide Mode
” on
page 71.
Turn the BpTRU off
if it is on. Hold
down the
Cycle
button while turning
the
BpTRU
on again.
Summary of Contents for BPM100
Page 2: ......
Page 14: ...Manuale di istruzioni BpTRU 8...
Page 20: ...BpTRU Operator s Manual 14...
Page 30: ...BpTRU Operator s Manual 24...
Page 48: ...BpTRU Operator s Manual 42...
Page 62: ...BpTRU Operator s Manual 56...
Page 82: ...BpTRU Operator s Manual 76...
Page 90: ...BpTRU Operator s Manual 84...