Vercise™ Neural Navigator 4 Programming Manual
Vercise™ Neural Navigator 4 Programming Manual
en
Placing ETS 3 into CP Mode (Pairing Mode)
1.
Press and hold the
Stimulation Button
on the side of the ETS until the lights flash (Figure 4). The ETS battery indicator light will
alternate blinking green and yellow, indicating that the ETS is in CP Mode (Pairing Mode) and is available for connection.
Note:
The Stimulator will automatically exit CP Mode (Pairing Mode) after two minutes if a CP connection is not established.
Figure 4.
ETS Battery Indicator Light
Connecting the Programming Wand to the CP (Vercise Gevia, Vercise PC, and ETS 2 Stimulators Only)
1. Connect the Programming Wand to the CP using the USB cable provided with the Programming Wand (Figure 5).
(a). Plug the Mini USB end of the USB Cable into the USB Port on the side of the Programming Wand.
(b). Plug the Standard USB end of the USB Cable into the USB Port on the CP.
Figure 5.
Vercise PC and Vercise Gevia: Clinician Programmer (CP) and Programming Wand
2. Wait for the Wand to perform a self‑test. At the end of the self‑test, the Wand will beep.
3. If the Power Light on the Wand is green, place the Wand over the Stimulator.
(a). If the Power Light on the Wand remains red, contact Technical Support.
Note:
Be sure to connect the Programming Wand to the CP prior to launching Vercise Neural Navigator.