6. POST-TREATMENT PROCEDURE
6.1. COMPLETING TREATMENT
1. Remove the probe from the endoscope.
Figure 53
Do not disconnect the probe and the handpiece
at this stage.
2. Switch off
IMPACT
and
ULTRASOUND
from the
READY
screen before starting this procedure.
3. Tilt the stone catcher.
Figure 54
4. Press the pedal halfway (STEP 1) for a few seconds
to empty the suction circuit and reduce the level of
water in the stone catcher.
1
Figure 55
To accelerate the emptying procedure, the stone
catcher can be disconnected from the handpiece.
5. The suction tubes must be cleared.
6. Loosen the probe from the handpiece, using one of
the following methods.
Figure 56