5. TREATMENT
Do not let the handpiece remain in contact with the
patient during treatment.
During treatment, an auditory information pulse will
be emitted.
This section provides guidance for using the product.
It does not provide detailed instructions for performing
lithotripsy procedures.
5.1. FUNCTIONAL TESTS
If a function or component is not working as
explained below, refer to the Troubleshooting
section.
1. From the
STAND BY
screen, press the
START
button to access the
READY
screen.
2. Insert the probe into a sterile receptacle of physio-
logical fluid.
3. Use the 2-mode foot pedal.
4. Press the pedal halfway (STEP 1) to activate suction
and make sure that suction is working properly (fluid
moving through the suction tube).
1
Figure 47
5. Press the pedal completely (STEP 2) to activate both
suction and energies and make sure that the quality
meter is in the green zone and the fluid is moving
through the suction tube.
1
2
Figure 48
6. Remove foot from the pedal to stop the functional
test.
Figure 49