30
7. Replace the console on a flat surface.
8. Refill the cooling system. Refer to the Filling the
Cooling System section.
8.2. REPLACING FUSES
1. Disconnect the power cord at the rear of the console.
Figure 74
2. Remove the fuse drawer located in the power socket.
Figure 75
3. Replace defective fuses with the fuse type specified
on the identification plate at the rear of the console.
4. Re-insert the fuse drawer.
5. If the fuses fail again, please contact your local Boston
Scientific sales representatives.
8.3. DOWNLOADING LOGFILE
Your local Boston Scientific sales representatives may
request this procedure.
1. Plug the USB key provided by EMS at the rear of the
console.
2. From the
STANDBY
screen, select
PARAMETERS
3. Press
LOGFILE DOWNLOAD
.
4. Follow the procedure displayed on the screen.
Summary of Contents for Swiss LithoCast Trilogy FT-235