Instructions for Use
Implantable Pulse Generator
en
Connecting the Lead, Lead Extension, Splitter, or Connector to the 16 Contact IPG
Note:
Lead Extensions, Adapters, and Splitters are not full-body MRI scan eligible.
IPG ports are labeled as follows:
C
D
Connecting to the IPG
•
Leads connect to IPG ports C or D.
•
For the Infinion 16, connect the Splitter tail with laser-etched bands (Contacts 1-8 of the Infinion 16
Lead) to the left Port C and the unmarked Splitter tail (Contacts 9-16 of the Infinion 16 Lead) to the
right Port D.
•
For the Infinion CX Lead, connect the tail with the single marker band (Contacts 1-8 of the Infinion
CX Lead) to the Port C and the tail with two marker bands (Contacts 9-16 of the Infinion CX Lead
to the right Port D.
•
For the Artisan and Artisan MRI 2x8 Surgical Leads, connect the left side to the left Port C. Connect
the right side (the laser-etched tail), Contacts 9-16, to the right port D.
Summary of Contents for SC-1216
Page 19: ...Implantable Pulse Generator 92469342 02 15 of 353 en This page intentionally left blank ...
Page 73: ...Implanteerbare pulsgenerator 92469342 02 69 van 353 nl Deze pagina is met opzet leeg gelaten ...
Page 179: ...Implanterbar impulsgenerator 92469342 02 175 af 353 da Denne side er med vilje efterladt tom ...
Page 213: ...Implanterbar pulsgenerator 92469342 02 209 av 353 sv Den här sidan har avsiktligt lämnats tom ...
Page 285: ...Wszczepialny generator impulsów 92469342 02 281 z 353 pl Strona celowo pozostawiona pusta ...
Page 357: ...Implantable Pulse Generator 92469342 02 353 of 353 This page intentionally left blank ...