Determining Longevity:
1. Fit the Precision system to patient,
make note of the amplitude, rate and
pulse width for each of the areas.
For multi-area stimulation please use
the Multiple-Area Battery-Longevity
Worksheet.
2. In the following chart, look up the
battery-longevity based on the ampli-
tude, rate and pulse width values that
correspond most closely to the settings
from step 1. (Interpolation may be used
for improved accuracy.) The battery-
longevity projections are given in
years
and represent the longevity of a typical
Precision battery when used continu-
ously. If the rechargeable Stimulator
battery is not used continuously, the
battery-longevity should be greater than
the value given.
Note:
• Cells in the table that are marked
with an “*” represent stimulation
settings that cannot be attained
with the Precision System.
Interpolation between a cell with
an “*” and another cell is not
possible.
• The longevity and recharge
interval estimates in this document
were generated using projections
of typical results based on Boston
Scientific’s internal testing data.
Actual results may vary.
Summary of Contents for Precision Spinal Cord Stimulator System
Page 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Page 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Page 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Page 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Page 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...