![Boston Scientific Precision Spinal Cord Stimulator System Clinical Manual Download Page 37](http://html.mh-extra.com/html/boston-scientific/precision-spinal-cord-stimulator-system/precision-spinal-cord-stimulator-system_clinical-manual_2796112037.webp)
Connecting to the IPG
Clinician Manual
en
Note:
If you experience difficulty when
inserting the lead, lead extension,
or port plug, use the hex wrench
to loosen (counterclockwise) the
setscrew and/or gently rotate the lead
to help advance the proximal end.
Note:
To confirm good connections, check
impedances before tightening the
setscrew.
2. Pass the hex wrench through the hole in
the septum located on the top of the IPG
header and tighten each set screw until
the hex wrench “clicks,” indicating lock.
CAUTION: Ensure that the lead is fully
inserted before tightening the setscrew
to prevent lead damage.
Note:
• If a port plug is used, it is still
necessary to tighten the setscrew on
the port plug, as described above.
• The hex wrench is torque-limited
and cannot be overtightened.
3. Place the IPG in the subcutaneous pocket
with label facing towards the skin.
4. Coil excess lead, extension or splitter
under the IPG.
CAUTION:
Do not excessively bend the
Splitter 2x8 receptacle. Bending the re-
ceptacle at a radius tighter than 2” (5 cm)
may result in loss of stimulation via one or
more contacts. Ensure the correct lengths
of leads are selected to prevent excessive
coiling.
5. Secure the IPG in the pocket by suturing
through the holes in the IPG header.
6. Close and dress the wound(s).
Summary of Contents for Precision Spinal Cord Stimulator System
Page 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Page 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Page 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Page 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Page 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...