![Boston Scientific Precision Spinal Cord Stimulator System Clinical Manual Download Page 48](http://html.mh-extra.com/html/boston-scientific/precision-spinal-cord-stimulator-system/precision-spinal-cord-stimulator-system_clinical-manual_2796112048.webp)
Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
90607862-30 42 of 309
en
from any Pulse Width Area screen. The
Rate screen is indicated by “Rate” in the
bottom left corner of the screen.
5. Area 1 will initially be highlighted. Press
the
"
button to cycle to the Area you wish
to adjust.
6. When the desired area screen is high-
lighted, press
p
or
q
to increase or
decrease the Rate.
7. To return to the main Level screen, press
and hold the
"
button for three seconds
until the Level screen appears.
Device Linking
A Remote Control can communicate with only
one Stimulator at a time. This prevents the
Remote Control from accidentally controlling
an unintended device. In the linking process,
the Remote Control will identify, by telemetry,
the intended Stimulator for communication.
1. Press and hold the
P
button to unlock the
Remote Control and initiate communica-
tion between the Remote Control and the
Stimulator.
2. If the Remote Control is not linked to a
Stimulator, it will display a “Link?” screen.
Press
P
to link. The Remote Control will
search for the nearest Stimulator.
If the Remote Control is already linked to
a Stimulator, it may continuously search
for it. This link must be cleared before a
new link can be established. (See “Clini-
cian Options” page 48).
3. The Remote Control will identify the
Stimulator by serial number. Verify before
proceeding.
4. Press
P
to confirm and continue. Press
"
to cancel.
When linking with a Trial Stimulator, the
Remote Control will ask whether existing
programs on the Trial Stimulator should
be cleared. If the programs on the Trial
Stimulator are cleared and if the Remote
Control contains existing programs, it will then
ask if these should be used.
Summary of Contents for Precision Spinal Cord Stimulator System
Page 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Page 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Page 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Page 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Page 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...