![Boston Scientific Precision Spinal Cord Stimulator System Clinical Manual Download Page 32](http://html.mh-extra.com/html/boston-scientific/precision-spinal-cord-stimulator-system/precision-spinal-cord-stimulator-system_clinical-manual_2796112032.webp)
Precision™ Spinal Cord Stimulator System Clinician Manual
Clinician Manual
90607862-30 26 of 309
en
Note:
• Ensure that the hex wrench is
fully seated in the setscrew before
tightening.
• The hex wrench is torque-limited
and cannot be overtightened.
4. Form an appropriately-sized pocket using
blunt dissection on either side of midline
for coiled excess lead and extension
connectors.
5. Place a small loop at the lead for slack.
If necessary, loosely tie a suture around
the lead-loop, but do not tighten onto the
lead.
CAUTION: Tightening sutures directly
on the lead can damage the lead.
6. Carefully remove excess slack by
gently pulling the extensions from the exit
wound.
7. For Permanent Trials, if desired, a small
suture may be used to close the exit
wound of the extension. Place and tape
a stress relief loop and dress the wound.
Proceed to the instructions for “Connect-
ing to the Trial Stimulator” page 32.
8. For Permanent Implant, close the midline
incision, and proceed to the instructions
Summary of Contents for Precision Spinal Cord Stimulator System
Page 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Page 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Page 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Page 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Page 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...