Permanent Lead Anchoring and Tunneling
Clinician Manual
en
Permanent Lead
Anchoring and Tunneling
Removing the Insertion Needle
1. Cut down around the insertion needle to
provide access for anchoring the lead.
2. Carefully withdraw the insertion needle
from the epidural space by slowly pulling
the needle up towards the proximal end of
the lead while holding the lead in place.
3. Once the insertion needle tip is exposed,
hold the lead as close to the exit site as
possible, then carefully pull the needle
completely from the lead.
Anchoring the Lead
Leads can be permanently anchored with a
suture sleeve or with a Clik™ Anchor. Refer
to the Clik Anchor Directions for Use, or
continue with the following steps to anchor
using a suture
sleeve
.
1. For percutaneous leads, carefully remove
the lead stylet using fluoroscopy to ensure
the lead position does not change.
2. Place a suture sleeve over the lead and
down to the supraspinous ligament or
deep fascial tissue.
3. Ligate the suture sleeve onto the lead by
tying a 2-0 silk or other nonabsorbable
suture around the center groove of the
sleeve to prevent sliding. Circumferential
stitches may be tied at the compression
slots.
CAUTION: Do not use polypropylene
sutures as they may damage the
suture sleeve. Do not suture directly
onto the lead or use a hemostat on the
lead body. This may damage the lead
insulation.
Note:
The 4 cm and 2.3 cm suture sleeves
each have three (3) compression
slots, which are designed to reduce
slippage.
4. Suture the sleeve to the supraspinous
ligament or deep fascia through the
suture sleeve holes.
5. Tie multiple sutures as tightly as possible
around the suture sleeve to secure it to
the lead.
CAUTION: Tightening sutures directly
on the lead can damage the lead.
6. For Permanent Trials, proceed to the
instructions for “Tunneling the Lead or
7. For Permanent IPG Implantation, proceed
to the instructions for “IPG Implantation”
page 34.
Summary of Contents for Precision Spinal Cord Stimulator System
Page 55: ...Rechargeable Stimulator System Clinician Manual 90607862 30 49 of 309 en ...
Page 77: ...Clinician Manual 90607862 30 71 of 309 en This Page Intentionally Left Blank ...
Page 131: ...Şarj Edilebilir Stimülatör Sistemi Klinisyen Kılavuzu 90607862 30 125 309 tr ...
Page 153: ...Klinisyen Kılavuzu 90607862 30 147 309 tr Bu sayfa özellikle boş bırakılmıştır ...
Page 291: ... نحشلا ةداعإل لباقلا هيبنتلا زاهج ماظن العالجي الدليل 309 من 285 90607862 30 ar ...
Page 315: ...Clinician Manual 90607862 30 309 of 309 This Page Intentionally Left Blank ...