BPAP System User Manual
6 / 57
Main device input
24 V, 3.33 A
Device offer to Heated Tubing Communications Port
24 V
18 W
Type of Protection against Electric Shock
Class II Equipment
Degree of Protection against Electric Shock
Type BF Applied Part
Degree of Protection against Ingress of Water
IP22
Pressure Range
Model
Work Mode
Pressure Range
G3 B20A
CPAP, S, AutoS
CPAP: 4.0
~
20.0 hPa
IPAP: 4.0
~
20.0 hPa
EPAP: 4.0
~
20.0 hPa
in 0.5 hPa increments
G3 B25S
CPAP, S
CPAP: 4.0
~
20.0 hPa
IPAP: 4.0
~
25.0 hPa
EPAP: 4.0
~
25.0 hPa
in 0.5 hPa increments
G3 B25A
CPAP, S, AutoS
G3 B25VT
CPAP, S, T, S/T
G3 B30SV
CPAP, S/T
CPAP: 4.0
~
20.0 hPa
IPAP: 4.0
~
30.0 hPa
EPAP: 4.0
~
25.0 hPa
in 0.5 hPa increments.
G3 B30VT
CPAP, S, T, S/T
G3 LAB
CPAP, AutoCPAP
S, AutoS, T, S/T
Under single fault conditions, ≤ 30 hPa for CPAP and AutoCPAP mode, ≤ 40 hPa for the rest
modes.
Pressure Display Accuracy
± (0.8 hPa
+
4%)
Static Pressure Stability
± 0.5 hPa
Ramp
The ramp time ranges from 0 to 60 minutes.
Sound Pressure Level
< 26 dB (A), when the device is working at the pressure of 10 hPa, Uncertainty: 2 dB (A).
Sound Power Level
< 34 dB (A), when the device is working at the pressure of 10 hPa, Uncertainty: 2 dB (A).
Summary of Contents for G3 B20A