BPAP System User Manual
7 / 57
Maximum Flow
Test of Maximum Flow rate for: G3 B25A, G3 B25S, G3 B25VT
Test Pressure
Pmin
Pmin + 1/4
(Pmax-Pmin)
Pmin + 1/2
(Pmax-Pmin)
Pmin + 3/4
(Pmax-Pmin) Pmax
Test Pressures (hPa)
4
10
15
20
25
Measured Pressure at the
Patient Connection Port (hPa)
3
9
14
19
24
Average Flow at the Patient
Connection Port (L/min)
90
150
150
150
150
When the working pressure is set to the values listed in the table, the average flow rate at
the patient end should be greater than 80% of the corresponding flow value in the table.
Test of Maximum Flow rate for: G3 B30VT, G3 B30SV, G3 LAB
Test Pressure
Pmin
Pmin + 1/4
(Pmax-Pmin)
Pmin + 1/2
(Pmax-Pmin)
Pmin + 3/4
(Pmax-Pmin) Pmax
Test Pressures (hPa)
4
11
17
24
30
Measured Pressure at the
Patient Connection Port (hPa)
3
10
16
23
29
Average Flow at the Patient
Connection Port (L/min)
90
150
150
150
120
When the working pressure is set to the values listed in the table, the average flow rate at
the patient end should be greater than 80% of the corresponding flow value in the table.
Test of Maximum Flow rate for: G3 B20A
Test Pressure
Pmin
Pmin + 1/4
(Pmax-Pmin)
Pmin + 1/2
(Pmax-Pmin)
Pmin + 3/4
(Pmax-Pmin) Pmax
Test Pressures (hPa)
4
8
12
16
20
Measured Pressure at the
Patient Connection Port (hPa)
3
7
11
15
19
Average Flow at the Patient
Connection Port (L/min)
85
135
140
140
140
When the working pressure is set to the values listed in the table, the average flow rate at
the patient end should be greater than 80% of the corresponding flow value in the table.
SpO
2
Range: 35%
~
100%
The margin of error for SpO
2
between 70% and 100% is ± 3%. No strict accuracy
requirements for SpO
2
below 70%.
Pulse Rate
Range: 30
~
240 BPM
Margin of Error: ± 2%