BPAP System User Manual
15 / 57
11. First Time Setup
11.1 Placing the Device
Place the device on a firm, flat surface.
WARNINGS!
• If the device has been dropped or mishandled, if the enclosure is broken, or if water enters
the enclosure, disconnect the power cord and discontinue use. Contact your home care
provider immediately.
• If the room temperature is above 95°F (35°C), the airflow generated by the device may
exceed 109.4°F (43°C). The room temperature must be kept below 95°F (35°C) while the
patient is using the device.
CAUTIONS!
• Always ensure that the device is placed in an area where the screen and indicators are
clearly visible.
• If the device has been exposed to very hot or very cold temperatures, allow it to acclimate
to room temperature (approximately 2 hours) before beginning setup.
• Make sure the device is away from any heating or cooling equipment (e.g., forced air vents,
radiators, air conditioners).
• The device is not suitable for use in high humidity environments. Make sure that no water
enters the device.
• Make sure that bedding, curtains, or other items are not blocking the filter or vents of the
device.
• Keep pets, pests or children away from the device and avoid small objects being inhaled or
swallowed.
• To avoid explosion, the device must not be used in the presence of flammable gases (e.g.
anesthetics).
• Tobacco smoke may cause tar to build-up in the device, which could lead to the
malfunctioning of the device.
• Air must flow freely around the device to allow it to function properly.
11.2 Installing the Air Filter and Filter Cap / PM2.5 Filter
(1) Attach the air filter to the filter cap, as shown in Fig. 11-1.
Filter Cap
Air Filter
Fig. 11-1