8.9
Dispensing
This input mode allows the user to dispense a set amount of fluid (in milliliters) at a set rate
(Motor % Speed).
Default settings:
Amount Per Dispense = 1,000 mL
Motor % Speed = 50%
To Enable Dispensing:
1
Open the
App Drawer
2
Select
Dispensing
3
Select
Settings
to configure
Dispensing settings
4
Set
Amount Per Dispense
(in mililiters)
5
Set
Motor % Speed
(0.04 -100)
6
Dispense Run Time will be
displayed.
Confirm by pressing
Save
Note: If your Dispense run time is shorter than 1 second the pump will generate a “Run Time
Too Short!” ERROR. Please reconfigure dispensing settings to be greater than 1 second
Page 28
®
FLEXFLO A4
8.0
Input Setup
Page 29
®
FLEXFLO A4
9.0
Output Setup
To Enable the Relay:
1
Open the
App Drawer
2
Select
Outputs
3
Select
Relay & Contacts
4
Set
Desired Values
(refer to chart below)
!
Contact #1
!
Contact #2
!
Contact #3
!
Relay Output
5
Confirm by pressing
Save
9.1
Relay
This feature is used to assign alarms to relay & contact closures
One value to be defined:
1) Relay Output
Selection:
Contact energizes when:
Pump Run/Stop
Motor turning (roller assembly is rotating)
Monitor Run/Fail
Motor fails to respond to commands
Manual Speed Active
Manual Speed mode is running
Prime Active
Prime mode is running
FVS
After the programmed delay time pulses are not received from flow sensor.
TFD
Tube failure is detected by sensors in the head
Both TFD/FVS
Either TFD or FVS system triggers