938461PK2/2-0822
9
8
Casting/Scanning a Locking Liner
Before you begin
Allow the user to wear the device for 10
minutes.
(See Section 11
.)
If a scanning method is used
Fit/trim the casting dummy to the size of the distal end of the liner, see 938413.
To reduce the amount of distal pressure within the final socket, the model should be lengthened
by 10-12
mm (depending upon redundant tissue etc.).
If a casting method is used
Fit/trim the casting dummy to the size of the distal end of the liner, see 938413.
We recommend using a casting method that accentuates weight bearing areas such as the
medial condylar flare in conjunction with applying tension to the distal pin; otherwise excessive
distal contact may result in the final socket, giving rise to oedematous swellings corresponding
to the distal perforations as shown below.
Additionally, when fabricating a diagnostic socket, we recommend a 5
mm pelite spacer is used
to further extend/lengthen the socket. When you are satisfied with the fit of the socket (by
adding any necessary socks) the spacer can, if necessary, be removed if excess distal contact
gives rise to distal oedematous swellings.
The softness of the silicone can mask the amount of distal contact that is occurring, this
will only be evident when oedematous swellings are observed.
The use of a diagnostic socket is strongly recommended to assess the amount of distal
pressure/contact that is occurring.
Oedematous
Swellings
Residual Limb
Distal End Surface
Distal End of Liner
(Inside view)
Oedematous
Swellings
938413
Summary of Contents for SBWTTCP22
Page 33: ...938461PK2 2 0822 33 Silcare Breathe Walk Silcare Breathe Walk 11 5 1...
Page 34: ...938461PK2 2 0822 34 2 Blatchford...
Page 35: ...938461PK2 2 0822 35 3 13 938413...
Page 36: ...938461PK2 2 0822 36 4 5...
Page 37: ...938461PK2 2 0822 37 5 1 1 pH 2 3 4 1 2 pH 3 4 5 6 7 30 C 5 2 1 2 3 5 3...
Page 39: ...938461PK2 2 0822 39 8 10 11 938413 10 12 938413 5 938413...
Page 40: ...938461PK2 2 0822 40 9 10 11 10 12 5...
Page 41: ...938461PK2 2 0822 41 1 2 3 10 10 4 5 6 12 10 11...
Page 42: ...1 2 3 4 938461PK2 2 0822 42 Loctite 243 3 Nm 1 8 1 4 12 13 5 19 13...
Page 93: ...938461PK2 2 0822 93 Silcare Breathe Walk Cushion Silcare Breathe Walk Locking Liner 11 5 1...
Page 94: ...938461PK2 2 0822 94 2 Blatchford...
Page 95: ...938461PK2 2 0822 95 3 13 938413...
Page 96: ...938461PK2 2 0822 96 4 Cushion Locking 5...
Page 97: ...938461PK2 2 0822 97 5 1 1 pH 2 3 4 1 2 pH 3 4 5 6 7 30 C 5 2 1 2 3 5 3...
Page 99: ...938461PK2 2 0822 99 8 10 11 938413 10 12 mm 938413 5 mm 938413...
Page 100: ...938461PK2 2 0822 100 9 10 11 10 12 mm 5 mm...
Page 101: ...938461PK2 2 0822 101 1 2 3 10 10 4 5 6 12 10 11...
Page 102: ...1 2 3 4 938461PK2 2 0822 102 Loctite 243 3 Nm 1 8 1 4 12 13 5 mm 19 mm 13...