©2022 Blackline Safety Corp
G7 Dock Technical User Manual
|
10
Rev. 11
2022-09-26
2.2
CUSTOMIZING INLETS
All four inlets are customizable, and a multi-lingual sticker sheet is provided to label
customized inlets.
Changes to your inlet configuration can be made in G7 Dock’s configuration profile, from the
Docks page on Blackline Live. The activation code on the back of the Dock can be used to
identify it from the list. If you are having difficulties accessing this page or finding G7 Dock in
Blackline Live, please contact our Customer Care team.
2.3
REACTIVE GASES
If you are using chlorine (Cl
2
) with G7 Dock, you must use the following:
•
Inlet 1, which has been optimized for Cl
2
and other highly reactive gases based on
order of gas calibration or bump testing.
•
Teflon lined tubing (ACC-FEP-T2).
•
A stainless steel regulator, or a regulator with a Viton valve seat and seals (ACC-DFR-V).
For more information on bump testing and calibrating with reactive gases, please see
Sensor Bump Testing and Calibration Order
2.4
CONNECTING G7 DOCK TO A CALIBRATION GAS CYLINDER
G7 Dock provides quick-connects that attach to one end of the tube. Join the tube to the
configured inlet at the back of G7 Dock by connecting both quick-connect ends and twisting
clockwise until it clicks securely into place. Ensure the other end of the tube is attached to a
demand flow regulator on the gas cylinder.