24
Chapter 2
December 2015
Flex Focus 1202 User Guide
(BB1756-Q)
HistoScanning
1
The scanner units to be used with HistoScanning™ systems from AMD have an extra
connector on the back of the monitor, in the upper left corner. It is used to take raw
data, before it has been converted to an image on the monitor, out of the system
.
To use the Flex Focus for HistoScanning, you must have a HistoScanning license
installed.
Figure 2-4. Connector for HistoScanning.
A cable (AO1385) is supplied to use with this connector.
Using the System with a Lithotriptor
Special considerations apply if you are using a lithotriptor with the system.
•
The lithotriptor must comply with the guidelines in EN 60601–1–1 [7]
(Safety
requirements for medical electrical systems).
•
You must have a license installed to use a lithotriptor with the system.
•
You must always follow instructions in the manufacturer’s user guide for the
lithotriptor.
Power Connections
Do
NOT
connect a lithotriptor to the auxiliary power outlet. Plug the system and the
lithotriptor directly into independent wall power outlets.
Connecting the System to the Lithotriptor
Connect the lithotriptor to a USB connector on the system. If necessary, use a USB
RS-232 adapter cable.
1. HistoScanning has not been market cleared by the FDA or licensed by Health Canada.
Pull firmly to open,
so you can attach
the connecting
cable
WARNING
LT-w1
To avoid an inaccurate location of the crosshair cursor due to misalignment of recorded
image and graphics, ensure the coordinate systems are synchronized.
Summary of Contents for Flex Focus 1202
Page 5: ...5 Chapter 6 Getting Started 59 Index 61 ...
Page 6: ...6 ...
Page 10: ...10 Chapter 1 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 44: ...44 Chapter 2 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 46: ...46 Chapter 3 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 58: ...58 Chapter 5 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 60: ...60 Chapter 6 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 64: ...64 ...
Page 65: ......
Page 66: ......