Flex Focus 1202 User Guide
(BB1756-Q)
Medical Equipment
Non-Medical Equipment
Power
source
WARNING
ME-w1
Follow the guidelines in EN 60601–1–1[7]
(Safety requirements for medical electrical
systems)
.
Equipment that complies with the requirements of EN/IEC 60601–1[2]
(Part 1: General
requirements for safety)
ANSI/AAMI ES60601-1 or CSA C22.2 No. 601.1–M90 [5] can be
connected to the system, but one of the following conditions must be fulfilled:
•
Each piece of equipment is plugged directly into an independent wall power outlet.
or
•
The system and other equipment are plugged into an external common isolation
transformer to control the leakage current during a ground connection fault.
If in doubt, contact your local BK Ultrasound representative.
Printers and
auxiliary
power outlet
ON/Standby
button does
not turn off
outlet
WARNING
ME-w2
An approved printer, specified in the Product Data sheet, can be connected to the internal
auxiliary power outlet on the system.
Do not use the auxiliary power outlet on this BK
Ultrasound system for any other equipment.
The auxiliary power outlet is still live even when the ON/Standby button on the system is
turned off. To remove voltage from this outlet, you must unplug the power supply cord
from the power source.
WARNING
NME-w1
Follow the guidelines in EN 60601–1–1[7]
(Safety requirements for medical electrical
systems)
.
If you connect non-medical equipment (instruments that do not comply with safety
requirements for medical equipment, such as a video monitor, video recorder, endoscopic
camera control unit or other documentation device), this equipment must be placed
outside the patient environment (1.5m from the bed, for example). The equipment must
fulfill the relevant EN standard or other applicable national or international standard.
One of the following conditions must be fulfilled:
•
The system and other equipment are plugged into an external common isolation
transformer to control the leakage current during a ground connection fault.
or
•
The system is grounded with an additional safety ground connection (see “Additional
Protective Ground and Potential Equalization” on page 18).
If in doubt, contact your local BK Ultrasound representative.
Summary of Contents for Flex Focus 1202
Page 5: ...5 Chapter 6 Getting Started 59 Index 61 ...
Page 6: ...6 ...
Page 10: ...10 Chapter 1 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 44: ...44 Chapter 2 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 46: ...46 Chapter 3 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 58: ...58 Chapter 5 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 60: ...60 Chapter 6 December 2015 Flex Focus 1202 User Guide BB1756 Q ...
Page 64: ...64 ...
Page 65: ......
Page 66: ......