(16-01642-01)
NOTE:
If the image depth is set very low (to see tissue close to the transducer with
high magnification), the needle tip echo can be outside the displayed image area. To
see the needle tip in this case, zoom out so the full needle path is visible or pan the
image to the side (to keep the high magnification).
Annotations – Labels, Bodymarks and Arrows
Labels, bodymarks and arrows are activated when you tap
Label
.
Labels
Figure 9-2. Label selector.
Verify
puncture
guide type
number
WARNING
P-w2
Verify that the type number of the needle guide displayed on the monitor corresponds to
the needle guide that you are actually using. If the number is incorrect, the puncture line
on the monitor may not correspond to the true puncture path in the tissue.
Watch the
needle tip
WARNING
P-w4
The puncture line on the image is an indication of the expected needle path. The needle
tip echo should be monitored at all times so any deviation from the desired path can be
corrected.
Offset
changes
WARNING
P-w3
Changes you make to the offset of a programmable needle guide or brachy matrix will
affect ALL programmable needle guides and brachy matrixes. This could lead to incorrect
puncture lines or matrix positions for a different guide than the one you wanted to
change.
Summary of Contents for bkSpecto
Page 8: ...8 ...
Page 10: ...10 Chapter 1 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 36: ...36 Chapter 3 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 104: ...104Chapter 9 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 120: ...120Chapter 11 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 130: ...130Appendix B August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 156: ...156Appendix C August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 162: ...162 ...
Page 163: ......
Page 164: ......