bkSpecto Advanced User Guide
(16-01642-01)
141
In the
Brachy Matrix
setup window, you make setup changes to brachy matrices and
needle guides. You can:
•
Move a programmable brachy matrix or needle guide left, right, in or out.
•
Specify that a brachy ruler is displayed on a sagittal view (E14CL4b transducer)
•
Define your own brachy matrix and ruler.
To move the offset in either direction:
•
Select the offset you want from the drop-down menu and tap
Save
.
Use the trackpad and pointer to navigate the Needle Guide window.
To have a vertical line displayed at the 0 of the brachy ruler:
•
Check the
Display ruler on sagittal views
checkbox and tap
Save
.
User-Defined Matrices (including Brachy and Transperineal)
User-defined matrices are listed in the lower part of the
Brachy Matrix Setup
window (Fig C-16).
You have the following options for user-defined matrices:
Offset
changes
WARNING P-w3
Changes you make to the offset of a programmable puncture guide or brachy matrix will
affect ALL programmable puncture guides and brachy matrices. This could lead to
incorrect puncture lines or matrix positions for a different guide than the one you wanted
to change.
Verify user-
defined
guide
WARNING B-w2
If you create a user-defined matrix, it is your responsibility to verify that the matrix that
appears on the monitor corresponds to the physical matrix you are using.
Summary of Contents for bkSpecto
Page 8: ...8 ...
Page 10: ...10 Chapter 1 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 36: ...36 Chapter 3 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 104: ...104Chapter 9 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 120: ...120Chapter 11 August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 130: ...130Appendix B August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 156: ...156Appendix C August 2018 bkSpecto Advanced User Guide 16 01642 01 ...
Page 162: ...162 ...
Page 163: ......
Page 164: ......