![BIRD SignalHawk SH-36S-PC Operation Manual Download Page 53](http://html1.mh-extra.com/html/bird/signalhawk-sh-36s-pc/signalhawk-sh-36s-pc_operation-manual_2756528053.webp)
Settings
38
Ampt & Trace Menu
Figure
28
Example, Ampt & Trace Menu
Autoscale
Resizes the graph to fit the whole trace on the screen. This function will change
the reference and scale of a trace.
Reference
Sets the y-axis value at the top of the graph.
Scale
Sets the dB value of each partition of the graph on a scale from 1 to 15. The
graph is partitioned into 10 divisions, giving a set number of dB per division.
Note:
This soft key is not displayed if the Units of measure is in
Volt or Watt. See
Attenuation
Controls the built-in attenuator on the signal input. This reduces the amplitude
of a high-powered signal. The attenuation can be set to Auto, or at levels of 0,
10, 20, and 30 dB. The reported value of the signal is automatically corrected for
the selected attenuation.
Note:
When the Reference (See
) is set to
a higher setting, Attenuation will automatically set itself above this
setting.
Example -
Reference is set to 25 dBm, Attenuation automatically
sets itself to 30 dB and cannot be lowered.
Summary of Contents for SignalHawk SH-36S-PC
Page 26: ...SignalHawk PC and Rack Mount Models 11 Figure 7 Select DSP Device List ...
Page 67: ...Measurements 52 Figure 41 Example Channel Power Integration Bandwidth STEP 2 STEP 3 ...
Page 116: ...SignalHawk PC and Rack Mount Models 101 Figure 75 Menu Map BW Sweep Menu All Measurements ...
Page 117: ...Menu Maps 102 Figure 76 Menu Map Amplitude All Measurements ...
Page 118: ...SignalHawk PC and Rack Mount Models 103 Figure 77 Menu Map Measurement All Measurements ...
Page 119: ...Menu Maps 104 Figure 78 Menu Map Mark Limit Menu All Measurements ...
Page 120: ...SignalHawk PC and Rack Mount Models 105 Figure 79 Menu Map File Help All Measurements ...
Page 121: ...Menu Maps 106 Setup Function Menu Maps Figure 80 Map Setup Mode Main Screen ...