Fluorescence Liquid Tests |
73
•
Aspirate 150 μL from wells C2
–F2, and dispense into wells C3–F3. Mix the wells
using the pipette. Do not discard the tips.
•
Aspirate 150 μL from wells C3
–F3, and dispense into wells C4–F4. Mix the wells
using the pipette. Do not discard the tips.
•
Aspirate 150 μL from wells C4
–F4, and dispense into wells C5–F5. Mix the wells
using the pipette. Do not discard the tips.
•
Aspirate 150 μL from wells C5
–F5, and discard the tips.
Fluorescence Polarization (FP) Test
•
Pipette 200 µL of the (green) polarization buffer (BUF) into wells A6–H6.
•
Pipette 200 µL of the green high polarization reference (HPR) into wells A7–B7.
•
Pipette 200 µL of the green low polarization reference (LPR) into wells A8–H8.
1
2
3
4
5
6
7
8
9
10
11
12
A
BUF
HPR
LPR
B
BUF
HPR
LPR
C
BUF
LPR
D
BUF
LPR
E
BUF
LPR
F
BUF
LPR
G
BUF
LPR
H
BUF
LPR