Clean the Objectives |
49
Do not touch the filters with your bare fingers!
1.
Open the access door on the front of the instrument. Slide the filter cube out
of its compartment.
2.
Inspect the glass filters for speckled surfaces or a “halo” effect. This may
indicate deterioration due to moisture exposure over a long period of time.
If you have any concerns about the quality of the filters, contact your
BioTek representative.
3.
Using cotton balls or lens-cleaning tissue moistened with a small amount of
high-quality alcohol, clean each filter by lightly stroking its surface in one
direction.
4.
Use a magnifying glass to inspect the surface; remove any loose threads left
from the cotton ball.
5.
Replace the filter cube and close the door.
Clean the Objectives
Applies only to models with the imaging module.
The objectives used in the Cytation 5 should be cleaned when necessary using
optical-grade swabs or lens paper moistened with lens cleaning solution or deionized
water.
Do not rub the lens.
Materials
•
Air puffer
•
Tweezers
•
Magnifying glass
•
Lens cleaning tissue
•
Optical-grade swabs
•
Cleaning solvent
Recommended Cleaning Solvents
Non-recommended Cleaning
Solvents
HyperClean (hexamethyldisoloxan and ethanol),
available from Olympus
Methyl ethyl ketone (MEK)
Isopropyl alcohol, 70%/30% with deionized water
Dimethyl ketone (acetone)
Methyl alcohol, 70%/30% with deionized water
Ethyl alcohol, 70%/30% with deionized water
Cytation 5
Summary of Contents for Cytation 5
Page 13: ...Installation ...
Page 31: ...14 Install the Imager Module 19 Cytation 5 ...
Page 37: ...Getting Started ...
Page 54: ...42 Getting Started BioTek Instruments Inc ...
Page 55: ...Maintenance ...
Page 66: ...54 Maintenance BioTek Instruments Inc ...
Page 67: ...Instrument Testing ...
Page 113: ...Specifications ...