18
| Installation
If you are installing an objective with a correction collar (i.e., a 20X, 40X, or
60X objective), be sure to grasp the objective by the adapter, not by the
correction collar, to avoid changing the correction collar settings.
2.
Screw each objective into its defined position. Do not overtighten the
objectives.
3c. Install the LED Cubes and Imaging Filter Cubes
1.
With the instrument powered off, slide the filter slide out of the instrument.
2.
Place the new LED cube in the appropriate position (Position 1, 2, 3, or 4) on
the filter slide.
3.
Use the 3/32" hex wrench to screw the LED cube onto the filter slide.
4.
Place the imaging filter cube on top of the LED cube you installed, and use
the hex wrench to screw it to the LED cube.
5.
Insert the LED cube’s wire clip into the socket on the carrier.
BioTek Instruments, Inc.