Luminescence Specifications |
109
Luminescence Specifications
The Cytation 5 measures luminescence from the top of 6- to 384-well plates. The
following requirements apply to 96-well plates with 200 µL/well, at room
temperature. Production testing is performed using a Harta plate.
Luminescence
DL
≤ 75 amol/well, 30 amol typical with low
-noise PMT
≤ 500 amol/well with red
-shifted PMT
Integration Time
10 seconds
Gain
150
Blank Wells
16
Imaging Specifications
The Cytation 5 imaging specifications are based on using an NIH 3T3 plate: 10,000
wells per cell, GFP stain, Costar 3603 black-sided, plastic-bottom plate.
Read Speed
At 20X, Autofocus on, 96-well plate, 1 image per well
< 10 minutes