56 | General Specifications
Absorbance Resolution
0.001 OD when operated in standalone mode
0.0001 OD when operated with Gen5
Accuracy
Normal Read Mode (96-well):
±1.0% ±0.010 OD from 0.000 to 2.000 OD @ 405 nm
Rapid Read Mode (96-well)
±2.0% ±0.020 OD from 0.000 to 2.000 OD @ 405 nm
Normal Read Mode (384-well):
±2.0% ±0.020 OD from 0.000 to 2.000 OD @ 405 nm
Rapid Read Mode (384-well):
±2.5% ±0.020 OD from 0.000 to 2.000 OD @ 405 nm
Linearity
Normal Read Mode (96-well):
±1.0% ± 0.010 OD from 0.000 to 2.000 OD at 405 nm
±3.0% ±0.010 OD from 2.000 to 3.000 OD @ 450 nm
Rapid Read Mode (96-well):
±2.0% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Normal Read Mode (384-well):
±2.5% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Rapid Read Model (384-well):
±2.5% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD)
Normal Read Mode (96-well):
±0.5% ± 0.005 OD from 0.000 to 2.000 OD @ 405 nm
Rapid Read Mode (96-well):
±1.0% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Normal Read Mode (384-well):
±1.5% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Rapid Read Mode (384-well):
±2.0% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
Throughput
From carrier start to carrier stop:
96-Well, Single Wavelength, Normal Read Mode: 38
seconds 96-Well, Single Wavelength, Rapid Read Mode:
26 seconds
384-Well, Dual Wavelength, Normal Read Mode: 3 minutes, 40
seconds 384-Well, Dual Wavelength, Rapid Read Mode: 2 minutes
Summary of Contents for 800 TS Series
Page 1: ...800 TS Absorbance Reader Instructions for Use BioTek Instruments Inc 2017 PN 1561011 Rev A ...
Page 12: ...xvi Safety Symbols BioTek Instruments Inc ...
Page 13: ...Installation ...
Page 28: ...BioTek Instruments Inc 24 Repackaging and Shipping Instructions ...
Page 29: ...Getting Started ...
Page 41: ...Maintenance ...
Page 48: ...BioTek Instruments Inc 44 Filter Storage and Handling ...
Page 49: ...Instrument Testing ...
Page 55: ...Specifications ...