30 | Operate the Reader Using the Touchscreen
BioTek Instruments, Inc.
The left side of the Main Menu screen lists the assay protocols defined on the reader. As
shipped, the reader contains no protocols and the list is empty. Up to 40 uniquely named
protocols can be programmed and stored. Tap
Next
to scroll through the list.
On the right side of the Main Menu are the following options:
•
Quick:
Define and run a simple single- or dual-wavelength protocol. You can
select the primary and secondary wavelength values and the plate type. If
applicable to your reader model, shake
and/or
incubate options are available.
•
Protocol:
Edit, create (and save), delete, and copy protocols. You can define
the protocol name; select the primary and secondary wavelength values, read
speed, and plate type; and define blank well(s). If applicable to your reader
mode, shake options are available.
•
Results:
View and optionally
print/export
measurement data stored on the
reader (for the 12 most recently run protocols).
•
Instrument:
Configure
the reader, printer, and USB flash drive settings; define
results output criteria; run an Absorbance Plate Test; and more.
Configure Your 800 TS
After you install the reader, and before you use it to create and run protocols, perform the
tasks in this section to define important instrument settings.
Set the Time and Date
When you turn on the 800 TS for the first time, set or confirm the date and time at your
location.
800TSI and 800TSUVI models:
When the incubator is turned on, the reader’s internal
temperature is displayed at the top of the screen, as shown above. When the incubator
is off, dashes are displayed.
The reader automatically performs delta OD and blank subtraction. To select the
dataset(s) for print or export, tap
Instrument
and then the
Output
tab.
Summary of Contents for 800 TS Series
Page 1: ...800 TS Absorbance Reader Instructions for Use BioTek Instruments Inc 2017 PN 1561011 Rev A ...
Page 12: ...xvi Safety Symbols BioTek Instruments Inc ...
Page 13: ...Installation ...
Page 28: ...BioTek Instruments Inc 24 Repackaging and Shipping Instructions ...
Page 29: ...Getting Started ...
Page 41: ...Maintenance ...
Page 48: ...BioTek Instruments Inc 44 Filter Storage and Handling ...
Page 49: ...Instrument Testing ...
Page 55: ...Specifications ...