If the examinee’s feet or hands are too dry, or has dead, hard skin built up, InBody720 may prompt the user to re-test the examinee. In this
case, wet the palms and soles with electrolyte tissue that comes with the InBody720 and re-test the examinee.
When wetting hands and feet, keep the moisture on hands and feet from dripping to the electrodes. Excessive moisture may cause an erosion
of electrode, which in turn may result in breakdown of the equipment. The electrolyte tissues provided by the Biospace are specifically
manufactured for InBody and thus is completely different from the generic wet tissues in the market. It is strongly recommended to use the
electrolyte tissues specifically designed for InBody720.
➌
Body Posture
During the test, the LCD monitors displays information on body composition, allowing the examinee to read the analysis
from their stance. When the examinee steps down from the stand, the InBody720 goes back to the home screen.
(1) Do not leave the arms by your side. Form an angle of 15 degrees between the arms and your side.
(2) Stand comfortably during the testing. Do not flex your muscles.
Disabled people who find standing for minutes a little bit difficult can get support from the back or side. In this case, there should be no
skin-to-skin contact between the supporter and the examinee. Testing is impossible with an amputee who has a thumb, an arm or a leg
amputated.