13
4. Transportation
It is recommended not to move the equipment once it is installed. If it is inevitable to relocate the equipment, follow the
safety requirements that follow.
(1) Turn the power off and remove plug from the power outlet before moving.
(2) Take all the possible measures to ensure no physical impact is made on the hand electrode.
(3) Adjust the level of the equipment using the legs of the equipment after moving.
(4) Tighten the connection between the lower and upper part with the hexagonal wrench.
A. Environmental Requirements
(1) Optimum Temperature: 20
℃
~70
℃
(2) Relative Humidity: 10%~95% (No condensation)
(3) Optimum Pressure: 50kPa~106kPa
B. Transporting Before Installation
Before installation, the equipment is in a packaging box provided by Biospace. Use a carrier to move the box over or have
two people hold both sides for safe transportation.
C. Transporting after installation
It is not recommended to move the equipment installed by Biospace or the authorized distributors of Biospace. If it is
inevitable to move the equipment, repack the equipment with the box and wrapping material the equipment came in, to
keep the equipment from being damaged during transportation.
After moving the equipment, adjust the level of the equipment using the bubble level indicator and legs of the equipment. The level of the
equipment is crucial to accurate testing.
Be careful with fragile freight. The package has fragile operation parts including LCD, which has the sign on the box.
After relocating the InBody720, make sure it is level again. Inaccurate leveling will affect accuracy of individual weight measurements.