56
1. Error Messages
The InBody720 displays the following error messages to warn the user of the problems run into during operations and to
guide the users to take steps. The following are the most common error messages and the steps to handle the
corresponding errors.
A. “After removing any objects on InBody720, press “Enter” button.
This message comes up when weight is detected from the base frame between
the power-on and the completion of boot-up process. Remove the object from
the base frame and restart the equipment.
B. “Enter personal data correctly.”
This message appears when the value for age or height of the examinee is out of the
permissible range for these data. Check your entry again. As for the permissible
range of each data, refer to the “Chapter 2, section 5 : Personal Profile.”
C. “Wipe hands and feet using electrolyte tissue.”
This message appears when the posture of the examinee is not appropriate or the
examinee’s palms or soles are too dry making it impossible to start the test. Correct
the posture of examinee or wet his/her palms and soles with electrolyte tissues,
before reinitiating the test.