SS69L DISSOLVED OXYGEN PROBE TRANSDUCER
SS69L Components
Order probe only as
RXPROBE02
Order interface only as
BSL-TCI16
The SS69L transducer measures dissolved oxygen. The SS69L includes a dissolved oxygen probe and an
interface to the BSL MP36/MP35 or MP30 unit.
See BSL
PRO
Lesson #A07 Fish Respiration and Q10.
Components
Dissolved
O
2
probe
Sodium Sulfate calibration standard (2.0 M Na
2
SO
3
)
Replacement membrane cap Dissolved O
2
electrode filling solution
Calibration bottle & pipette
Polishing strips
Interface
Use with BIOPAC BSL-TCI16 Transducer Connector to record with a BIOPAC data acquisition
unit.
Usage
There are four steps to using the Dissolved O
2
probe:
1.
Setup
2.
Warm-up
3.
Calibration
—
optional
4.
Recording
1.
Setup
a.
Remove and discard the blue protective cap from the tip of the
probe.
b.
Unscrew the membrane cap from the tip of the probe.
c.
Use a pipette to fill the membrane cap with 1 mL of the Electrode
Filling Solution.
d.
Carefully thread the membrane cap back onto the electrode.
e.
Place the probe into a beaker filled with about 100 mL of distilled
water.
2.
Warm-up
a.
Insert the BT connector on the RXPROBEO2 into the BSL-TCI16
transducer connector.
b.
Connect the BSL-TCI16 transducer connector to the MP data acquisition unit.
c.
Turn the MP unit ON and wait 10 minutes for the probe to warm up.
The probe must stay connected to the interface at all times to keep it warmed up. If the probe
is disconnected for more than a few minutes, the warm-up routine will need to be repeated.