![BIOPAC Systems MP35 Hardware Manual Download Page 124](http://html1.mh-extra.com/html/biopac-systems/mp35/mp35_hardware-manual_2754784124.webp)
www.biopac.com
BIOPAC Hardware |
GASSYS2
| Page 2 - 4
Updated: 8.7.2013
Dimensions/Weight
1.7-liter chamber
5-liter chamber
Length:
22 cm
55 cm
Height:
18 cm
18 cm
Width:
14 cm
14 cm
Weight:
1.38 kg
1.96 kg
GASSYS2 Software Scaling Defaults
Acq
Knowledge
Software
Default %O
2
Scaling
Default %CO
2
Scaling
BSL
PRO
Software
Default %O
2
Scaling
Default %CO
2
Scaling
Calibration Procedure
The GASSYS2 scaling parameters are factory-set to a “ballpark” default, but in order to achieve accurate
measurements, these values must be re-calibrated in the Acq
Knowledge
or BSL
PRO
software using the
following steps.
1.
Flood the mixing chamber with fresh (ambient) air. (20.93% O
2
, 0.04% CO
2
) This can be accomplished
by attaching the calibration syringe to the mouthpiece and cycling fresh air into the mixing chamber.
2.
Monitor the gas concentration levels using Acq
Knowledge
or BSL
PRO
software. When the levels appear
stabilized at ambient, obtain the first calibration point. (CAL 1)
3.
Flood the chamber with the gas mixture using BIOPAC’s GASCAL and GASREG. The recommended
concentration is 16% O
2
, 4% CO
2
and 80% nitrogen (N
2
).
4.
When the levels appear stabilized, obtain the second calibration point. (CAL 2)