www.biopac.com
BIOPAC Hardware |
ITBS Circulator
| Page 1 - 3
Updated: 2.6.2013
CIRCULATOR A/B HEATING CIRCULATORS
Heating circulators are used
with Tissue Bath Stations and
include a digital temperature
display and the following
controls:
Preset
Temperature
Power
Heater
Circulation
Inlet
and
Outlet
ports are on
the back, along with the
power cord.
Circulator A
:
110 V, 60 Hz
Circulator B
:
220 V, 50 Hz
CIRCULATOR SETUP AND USAGE GUIDE
BIOPAC Heating Circulators will maintain
water temperature at a preset value in the
range 30
0
C to 45
0
C and circulate the water
through tissue baths.
Heating circulators include a digital temperature
display and the following controls:
Preset
Temperature
Power
Heater
Circulation
CALIBRATION
Although the offset value for the temperature sensor
is factory-calibrated, the user can calibrate the
controller’s internal temperature sensor. To calibrate
the sensor:
1.
Install a calibrated reference thermometer in
the bath.
2.
Adjust the offset value to zero.
3.
Adjust the preset value to an appropriate
temperature.
4.
Once the bath reaches the preset value and
stabilizes, calculate the offset value by
noting the difference between the reference
thermometer value and the preset value.
5.
Enter this value as an offset.
ERROR CODES
Display
Indication
Lo
Water in the bath is not enough or
the bath is empty.
Sen
Microprocessor
cannot
communicate with the temperature
sensor.