TROUBLESHOOTING
There is no water circulation or very little.
1.
Check the hose connections and be sure they are connected to the correct positions.
2.
Check that the hoses are not bent or twisted (which might impede the flow of water).
3.
Confirm that there is at least 4 liters of water in the bath.
There is some air in the waterway.
To remove the air:
1.
Press the PUMP ON/OFF key to
OFF
stop the circulator pump.
2.
Disconnect the hose from the INPUT of tissue bath. (Leave other end connected to the Circulator
OUTLET.)
3.
Put the end of the hose in a bucket to catch the water flow.
4.
Press the PUMP ON/OFF to
ON
to start the circulator pump.
5.
Operate the circulator pump for a few 1-2 second cycles.
6.
Press the PUMP ON/OFF key to
OFF
stop the circulator pump.
7.
Reconnect the hose to the INPUT of the tissue bath.
8.
Press the PUMP ON/OFF to
ON
to start the circulator pump and continue with normal operation.
TECHNICAL SPECIFICATIONS
Temperature Range:
30
0
C to 44
0
C
Reading Sensitivity:
0.1
0
C
Display:
3 digit (LED Display)
Water Bath Volume:
4.5 liters (Stainless Steel)
Circulation Flow:
2 liter/min.
Heater Resistance:
1000 Watt
Circulation Pump:
110 V 100W Plastic Head
Supply Voltage:
CIRCULATA:
110 Volt 60 Hz (1000 Watt)
CIRCULATB:
220 V 50 Hz (1100 Watt)
Inlet/Outlet
OD 8.5mm, ID 6.3mm Tubing
Temperature Offset Range:
0
0
C to 1.2
0
C
Acceleration Levels:
0 to 5
HARDWARE GUIDE