Irys® User Guide
07/10/2015
23
For Research Use Only. Not Intended For Diagnostic Purposes.
2. Then using another clean tissue, put a few drops of Sparkle on the tissue and following the same process to
wipe the Objective, clean with Sparkle to remove any excess oil. Dry the Objective with new clean tissue.
3. Using the Zeiss oil and paddle, dispense one drop of oil onto the Objective lens.
4. If a bubble forms when oil is dispensed, wipe off the Objective, clean with Sparkle, and repeat the application of
the oil.
Note:
Do not apply excessive oil to the Objective. Enough oil should be applied to form a rounded droplet that
does not run down the side of the lens.
5. Select the Oil Objective checkbox when finished oiling the Objective.
Figure 17: Load Sample Screen Complete
Forward
1. After all required tasks are complete, press the Forward button to proceed to the Focus & Register screen.
2. The Status icon will change from the green checkmark to the transparent hourglass timer to indicate a process
is in action.
Note:
There are intermediate steps described below, which alert the user if the loading is out of specification.
Sample Detect
1. When the XY Stage retracts, Irys will check the IrysChip cartridge for a present sample.
2. If a sample is not detected, Irys will notify the user and ask to fix the problem or continue with the workflow.
3. If the user selects to fix the problem, Irys will eject the XY Stage to have the sample loaded.