43
5061-049 REV 1.4
25-9574-75UM AE 05/2010
43
Step 4
To enter the Results Screen with the selected parameters press OK
OR
Cancel the selections and return to the Applications Screen by pressing Cancel
Esc
Step 5
Pipette on the reference sample and lower the sampling head. If Auto-Read is
off, press the 0A/100%T key. This will be used for all subsequent samples until
changed. If QA is switched on the sample will need to be replaced and the
0A/100% T key pressed again.
Step 6
Clean the top and bottom plates, pipette on the sample and lower the sampling
head. If Auto-Read is off, press
.
Repeat step 6 for all samples.
Results Screen
The Result at the selected wavelength is displayed on the screen.
Use the left and right arrows to move the cursor and display the value at the
cursor position (+/- 15 nm from set wavelength).
Press Cancel
Esc
to return to the Applications Screen.
Press
to display available Options which are described below.
Options (select using keypad numbers)
1. Return to Parameters Screen (step 1 above).
2. Print result via selected method.
3. Toggle between Absorbance and %T mode.
4. Print Graph – greyed out if no data are available.
7. Sample Number – add a prefix to the sample number and reset the
incrementing number to the desired value.
8. Save Method – use the left and right arrows to select a folder to store in
(Favourites/Methods 1–9), press the down arrow and enter name.
9. Auto-Print – toggles Auto-Print on/off.
Exit Options by pressing
Esc
, or wait.