8.3.2 Starter replacement (P/N 100-1370)
•
Turn off the power.
•
Disconnect the power cord.
•
Remove the four screws located on the side of the instrument.
•
Remove the lid.
•
Place it down on its backside on a non-abrasive surface.
Caution: Do not put the UV filter directly on the bench.
•
To replace the starter, rotate it counter-clockwise and remove.
•
To replace the starter insert it into the holder and rotate clockwise.
•
Reassemble the lid and retighten the screws on both sides.