4.2.2 Assembling the CCD camera
Gel Doc 2000 RS-170/CCIR Camera (170-7951/170-7952)
a) Carefully take the CCD camera assembly out of the box and hold the camera so that the two
locking holes are facing you.
b) Place one plastic black screw into the rear flat side of the MZL.
c) Insert the CCD camera into the bracket from the open left side and fix it with one or two
black screws. Place the bracket with the camera on the top of the Hood.
d) Verify that the MZL is adherent to the black gasket and fix the bracket with the four washers
and knobs.
e) A completely assembled camera will look like the photo shown below:
Gel Doc CCD Camera assembled on the bracket
Black plastic screw
required for proper
light tight operation
Please refer to the instructions included in
Appendix A
for a complete description of the cables
connection.
ChemiDoc RS-170/CCIR Camera: (170-7953/170-7954)
a) Carefully take the CCD camera assembly out of the box and hold the camera so that the two
locking holes are facing you.
b) Follow the above instructions (b-d).
c) A completely assembled camera will look like the photo shown below:
ChemiDoc CCD Camera assembled on the bracket