Focusing Ring
Iris Control Ring
12. When using this Lens for most sensitivity in Chemi Imaging, make sure that the Iris is opened
all the way.
4.3 Description of Functions and System Initialization
4.3.1 Control panel
The front membrane touch pad control panel of the Universal Hood allows full control of all the
functions including the lens operations, the UV, Epi and White Light (optional) illumination in
the darkroom cabinet. Please read the following section to become familiarized with each
function.
Imaging Indicator LED
Power Indicator
LED
Prep UV LED and
button
Trans UV LED and
button
Hold LED and
button
Trans White LED
and button
Epi-White LED and
button
Membrane touch pad – Light Source Section
The
Power
light tells you whether the system is turned on or not. The power on/off switch is
located on the back left side of the instrument.
The
Imaging
led blinks when you are acquiring an image with the ChemiDoc camera (this
function is not present using the Gel Doc 2000 or ChemiDoc XRS cameras) in the integration
mode. It also blinks if the computer is turned off.
Page 19 of 42